使用TDLAS系统的低浓度CO检测
作者:

Low concentration CO detection using the TDLAS system
Author:
  • 摘要
  • | |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • | |
  • 文章评论
    摘要:

    目前基于可调谐半导体激光吸收光谱(TDLAS)技术,已经可以实现许多种类气体浓度的高精度检测。但在实际的工程应用和仪器的批量生产当中,囿于一些激光器本身性能以及检测仪器体积的限制(无法使用更大的反射池),以及环境因素的影响,给低浓度气体分辨造成困难或者使检测精度受到影响。为改善这一问题,本文将主要介绍一种采用固定波门和差值的计算方法,来实现对低浓度的一氧化碳有效分辨,并在不同的温度条件下,对低浓度一氧化碳进行实测。检测结果表明,在恒定的温度条件下,检测1001ppm的CO的误差为22,在12℃的温差范围内,检测最大误差为-172ppm。在不同的环境条件下,采用这种处理方式,能够有效地进行低浓度一氧化碳检测,测试结果能够满足工程实际使用需求,一定程度上抵消器件性能造成的影响。在批量较大或器件存在较大个体差异时,本文对于仪器研制设计和生产制造具有积极作用。

    Abstract:

    At present,high precision detection of the concentration of many types of gases can be achievedbased on the tunable semiconductor laser absorption spectroscopy(TDLAS)technology.However,in actual engineering applications and mass production of instruments,the performance of some lasers themselves and the limitations of the volume of the detection instrument(larger reflector cells cannot be used),as well as the influence of environmental factors,make it difficult to distinguish low concentration gases or affect the detection accuracy.To improve this problem,a calculation method using a fixed wave gate and a difference is mainly introduced to achieve effective resolution of low concentration carbon monoxide,and low concentration carbon monoxide are measuredunder different temperature conditions.The test results show that the error of detecting 10.01 ppm CO is 2.2%under constant temperature conditions,and the maximum detection error is 1.72 ppm within the temperature difference range of 12℃.Under different environmental conditions,the processing method is able to effectively carry out the detection of low concentration carbon monoxide,and the test results can satisfy the requirements of the actual use of the project,offsetting the impact of device performance to a certain extent.This paper has a positive effect on instrument development,design and production when the batch is large or there are large individual differences in the devices.

    参考文献
    相似文献
    引证文献
引用本文

王国水,张超峰,栗佩康,裴彦军,李彬,朱旺旺,赵明阳.使用TDLAS系统的低浓度CO检测[J].激光与红外,2025,55(3):368~373
WANG Guo-shui, ZHANG Chao-feng, LI Pei-kang, PEI Yan-jun, LI Bin, ZHU Wang-wang, ZHAO Ming-yang. Low concentration CO detection using the TDLAS system[J]. LASER & INFRARED,2025,55(3):368~373

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 在线发布日期: 2025-03-14